User manual TEKTRONIX 2213 OPERATORS

DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!

If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual TEKTRONIX 2213. We hope that this TEKTRONIX 2213 user guide will be useful to you.


TEKTRONIX 2213 OPERATORS: Download the complete user guide (949 Ko)

You may also download the following manuals related to this product:

   TEKTRONIX 2213 SERVICE (14865 ko)

Manual abstract: user guide TEKTRONIX 2213OPERATORS

Detailed instructions for use are in the User's Guide.

[. . . ] COMMITTED TO EXCELLENCE PLEASE CHECK FOR CHANGE INFORMATION AT THE REAR OF THIS MANUAL. 2213 OSCILLOSCOPE OPERATORS INSTRUCTION Tektronix, Inc. Box 500 Beaverton, Oregon 070-3397-00 Product Group 46 97077 MANUAL Serial Number __________ First Printing JUN 1981 Revised AUG 1982 2213 Operators CONTROLS, CONNECTORS, AND INDICATORS The following descriptions are intended to familiarize the operator with the location, operation, and function of the instrument' controls, connectors, and indicators. s be maintained as changes occur in the intensity level of the trace. POWER, DISPLAY, AND PROBE ADJUST 4 0 1 0 Refer to Figure 3 for location of items 1 through 7. Internal Graticule-Eliminates parallax viewing error between the trace and graticule lines. [. . . ] Set A SEC/DIV to 10 I_LS, and A & B INT to CH 1 or CH 2 as appropriate for applied signal. Set A TRIGGER SLOPE for a positive-going signal (lever up) if the applied TV signal sync pulses are positivegoing, or for a negative-going signal (lever down) if the TV sync pulses are negative-going. OPERATOR' ADJUSTMENTS S INTRODUCTION Two adjustments should be performed before making measurements with your oscilloscope: Trace Rotation and Probe Compensation. Before proceeding with the following adjustment instructions, verify that the correct line fuse is installed (refer to the " Preparation for Use" information). Verify that the POWER switch is OFF (button out), then plug the power cord into the ac-power-input source. Push in the POWER switch (ON) and allow a 20-minute warm-up time before starting these adjustments. NOTE Normally, the resulting trace will be parallel to the center horizon tal graticule line, and the Trace Rotation adjustment should not be required. If the resulting trace is not parallel to the center horizontal graticule line, use a small flat-bit screwdriver to adjust the TRACE ROTATION control and align the trace with the center horizontal graticule line. PROBE COMPENSATION Misadjustment of probe compensation is one of the sources of measurement error. To ensure optimum measurement accuracy, always compensate the oscilloscope probe before making measurements. Pre set instrument controls and obta in a baseline er lnstrumen t Familiar izati on") . Use the Channel 1 POSITION control to move the baseline trace to the center horizontal graticule line. REV OCT 1981 15 2213 Operators 1. Preset instrument controls and obtain a baseline trace (refer to " Instrument Familiarization" ). CORRECT FLAT 2. Connect the two 10X probes (supplied with the instrument) to the CH 1 and CH 2 input connectors. Set both VOLTS/DIV switches to 0. 1 (10X PROBE) and set both AC-GND-DC switches to DC. Select CH 1 VERTICAL MODE and insert the tip of the Channel 1 probe in the PROBE ADJUST output jack. Using the approximately 1-kHz PROBE ADJUST square-wave signal as the input, obtain a display of the Instrument Familiarization" ). Set the SEC/DIV switch to display several cycles of the PROBE ADJUST signal. Check the waveform presentation for overshoot and rolloff (see Figure 10). If necessary, adjust the probe compensation for flat tops on the waveforms. Refer to the instructions supplied with the probe for details of compensation adjustment. Figure 10. Probe compensation. OVER COMPENSATED (OVERSHOOT) UNDER COMPENSATED (ROLLOFF) 465/DM-O-5 8. Select CH 2 VERTICAL MODE and connect the Channel 2 probe tip to the PROBE ADJUST output jack. Use the Channel 2 POSITION control to vertically center the display and repeat step 7 for the Channel 2 probe. BASIC APPLICATIONS After becoming familiar with all the capabilities of the 2213 Oscilloscope, the operator can then adopt a convenient method for making a particular measurement. [. . . ] Measured with a vertically centered 5-divisio reference signal from a 50-a n source driving a 50- a coaxial cable that is terminated in 50 a at the input connector, with the VOLTS/DlV Variable control in its CAL detent. Rise time is calculated from the formula: 0. 35 BW (in MHz) Step Response Rise Time Rise Time = Bandwidth Measured with a vertically centered 6-division reference signal from a 50-a source driving a 50- a coaxial cable that is terminated in 50 s2 , both at the input connector and at the P6120 probe input, with the VOLTS/DIV Variable control in its CAL detent. o" to +4o " c c 20 mV to 10 V per Division 2 mV to 10 mV per Division +4o " to +5o " c c 2 mV to 10 V per Division Chop Mode Repetition Rate aPerformance Requirement not checked in Service Manual. D C to at least 60 MHz. D C to at least 50 MHz. D C to at least 50 MHz. ~ 250 kHz +30%. @ 27 2213 Operators Table 2 (cant) Characteristics I Performance Requirements VERTICAL DEFLECTION SYSTEM (cont) Supplemental Information Input Characteristics Resistance Capacitance Maximum Safe Input Voltage DC Coupled 1 Ma+2% . a 30 pF It3 PF. ~ 1. A 400 V (dc + peak ac) or 800 V p-p ac to 1 kHz or less. a 400 V (dc + peak ac) or 800 V p-p ac to 1 kHz or less. a At least 10 to 1 at 10 MHz. [. . . ]

DISCLAIMER TO DOWNLOAD THE USER GUIDE TEKTRONIX 2213




Click on "Download the user Manual" at the end of this Contract if you accept its terms, the downloading of the manual TEKTRONIX 2213 will begin.

 

Copyright © 2015 - manualRetreiver - All Rights Reserved.
Designated trademarks and brands are the property of their respective owners.